InVivo Dosimetry detector table for selection of the appropiate detector.
| Photons | Electrons | Cobalt | |
|---|---|---|---|
| Entrance dose, few corrections | EDP-103G (4-8 MV) EDP-153G (6-12 MV) EDP-203G (10-20 MV) EDP-HL3G (16-25 MV)*1 |
EDD-23G EDP-53G |
EDP-53G (1.25 MeV) |
| Total body irradiation (TBI) | EDD-53G *2 | - | - |
| Entrance dose, low perturbation | EDD-23G EDP-HL3G (16-25 MV) |
EDD-23G | EDD-23G |
| Exit dose | EDD-23G Any detector can be used |
- | EDD-23G |
| Risk organ monitoring, measurements outside the field |
EDD-53G | EDD-53G | EDD-53G |
| Intracavitary (Teletherapy) | IDF-13G | - | - |
| Brachytherapy | IDF-13G (e.g.rectum) IDF-3 (e.g.rectum) IDF-5 (e.g.rectum) IDF-thin (e.g.bladder) |
||
*1 High energy, low perturbation; can be used up to 25 MV
*2 When using spoiler (build-up), otherwise same recommendations as entrance dose, few corrections