Check your complete Bragg peak depth dose curve distribution for single beam spot delivery
Giraffe is a high resolution multi-layer ionization chamber, which consists of 180 large 12 cm diameter electrodes.
Instant absolute range verification for weekly & monthly QA
3x faster commissioning of your particle therapy rooms
Reconstruct the Bragg peak for a single spot within seconds
Verify the clinical range of up to 230 MeV
Instant daily PBS verification within one second
- Check your complete Bragg peak depth dose curve distribution for single beam spot delivery
Minimize loss of data collection
- Save time with reduced effort to correct data at TPS level. Verify the TPS library periodically without needing to set up the water tank.
Technical specs
- Water equivalent material for equivalent scattering properties
- Measure ranges with an accuracy of 0.5 mm compared to measurements in water (Bortfeld fit)
- Verify the full clinical range with proton beam energies up to 230 MeV
180 chambers with 2 mm pitch
439 (L) x 17 (W) x 19.5 (H)